VOCs在线监测进样方式对非甲烷总烃测定的影响
Effects of Injection Methods of VOCs On-line Monitoring on the Determination of Non-methane Hydrocarbons
投稿时间:2019-08-27  修订日期:2020-05-12
DOI:10.19316/j.issn.1002-6002.2020.05.19
中文关键词:  VOCs在线监测系统  非甲烷总烃  大气平衡
英文关键词:on-line monitoring system of VOCs  non-methane hydrocarbons  atmospheric equilibrium
基金项目:
作者单位
葛明民 江苏天瑞仪器股份有限公司, 江苏 昆山 215300 
韩高锋 江苏天瑞仪器股份有限公司, 江苏 昆山 215300 
丁建峰 江苏天瑞仪器股份有限公司, 江苏 昆山 215300 
张冬冬 江苏天瑞仪器股份有限公司, 江苏 昆山 215300 
王小军 江苏天瑞仪器股份有限公司, 江苏 昆山 215300 
张苏伟 江苏天瑞仪器股份有限公司, 江苏 昆山 215300 
刘召贵* 江苏天瑞仪器股份有限公司, 江苏 昆山 215300 
通讯作者:刘召贵*  江苏天瑞仪器股份有限公司, 江苏 昆山 215300  
摘要点击次数: 147
全文下载次数: 4
中文摘要:
      采用正压、负压、大气平衡3种进样方式,研究了VOCs在线监测系统对非甲烷总烃测定的影响。探讨在大气平衡进样方式下定量环内压力、平衡时间、进样流量对非甲烷总烃测定的影响,研究结果表明:采用正压进样方式,定量环内正压增大,非甲烷总烃质量浓度偏高;采用负压进样方式,定量环内负压增大,非甲烷总烃质量浓度降低;采用大气平衡进样方式,定量环内压力、平衡时间、进样流量的增大或减小对非甲烷总烃测定没有明显影响。比较3种进样方式,大气平衡进样方式的VOCs在线监测系统能够提高非甲烷总烃在线监测结果的准确性和可靠性,更适用于污染源废气在线监测。
英文摘要:
      The effects of volatile organic compounds (VOCs) on-line monitoring system with positive pressure injection,negative pressure injection and atmospheric equilibrium injection of three different injection methods on monitoring the non-methane hydrocarbons (NMHC) were investigated in this paper.Under the atmospheric equilibrium injection mode,the impacts of the pressure of the sample loop,balance time and the flow rate of sample on the determination of NMHC were discussed.The results showed that the concentration of NMHC increased with the increasing sample positive pressure of the sample loop under the positive pressure injection mode,and the concentration of NMHC decreased with the increasing sample negative pressure of the sample loop under negative pressure injection mode.The pressure of sample loop,injection flow rate and balance time had no significant effect on the concentration of NMHC with atmospheric equilibrium injection.A comparison of three different injection methods indicated that VOCs on-line monitoring system with atmospheric equilibrium injection method could improve the accuracy and reliability of on-site monitoring results of NMHC,which might be more suitable for the on-line monitoring of stationary pollution source.
查看全文  查看/发表评论  下载PDF阅读器